310
Views
0
CrossRef citations to date
0
Altmetric
Research

Tear meniscus height agreement and reproducibility between two corneal topographers and spectral-domain optical coherence tomography

ORCID Icon, ORCID Icon, ORCID Icon & ORCID Icon
Received 24 Dec 2023, Accepted 06 Mar 2024, Published online: 16 Apr 2024

ABSTRACT

Clinical relevance

Tear meniscus height (TMH) is an important clinical marker in dry eye diagnosis and management.

Purpose

To evaluate the reproducibility and agreement of TMH measurements in non-clinical participants using the Oculus Keratograph 5 M, Medmont Meridia, and Spectral-domain optical coherence tomography (Spectralis SD-OCT).

Methods

Fifty-six participants (mean 43.8 ± 22.4 years) were recruited for this cross-sectional study. Image acquisitions were performed on the three devices, sequentially and randomized. The repeatability and reproducibility of inter-observer and inter-device analysis were performed. Repeated measures ANOVA and Bland-Altman Plots were used to evaluate the agreement between devices.

Results

The mean TMH with the Oculus Keratograph 5 M, Medmont Meridia and Spectralis SD-OCT were 0.29 ± 0.16 mm, 0.24 ± 0.09 mm and 0.27 ± 0.16 mm, respectively. There were no significant inter-observer differences (paired t-tests, p < 0.001). All the devices exhibited good inter-observer reliability (ICC ≥ 0.877), and good repeatability (CV ≤ 16.53%). Inter-device reliability is moderate (ICC = 0.621, p < 0.001). Repeated measures ANOVA revealed that TMH measurements given by the Spectralis SD-OCT are not significantly different from the Oculus Keratograph 5 M (p = 0.19) and the Medmont Meridia (p = 0.38). TMH measurements from Oculus Keratograph 5 M were significantly higher than those from Medmont Meridia (p = 0.02). Correlations between the mean TMH and the difference in the TMH measurements were positive for Oculus Keratograph 5 M and Medmont Meridia (r2 = 0.62, p < 0.001), negative for Medmont Meridia and Spectralis SD-OCT (r2 = -0.59, p < 0.001), and not significant for Oculus Keratograph 5 M and Spectralis SD-OCT (r2 = 0.05, p = 0.74). A strong correlation was found for TMH measured with all devices (r2 = 0.55 to 0.81, p < 0.001).

Conclusions

The Oculus Keratograph 5 M, Medmont Meridia, and Spectralis SD-OCT provide reliable and reproducible inter-observer TMH measurements. Inter-device reliability is moderate, with a close correlation between Spectralis SD-OCT and the Oculus Keratograph 5 M. Oculus Keratograph 5 M and Medmont Meridia are repeatable devices appropriate for the measurement of TMH, but they are not interchangeable in clinical practice.

Introduction

Dry eye disease (DED) is a common ophthalmic disorder,Citation1 with a 5% to 50% prevalence.Citation2 DED is characterised by a loss of tear film homoeostasis and hyperosmolarity, with symptoms and signs of tear film instability such as inflammation, ocular surface damage, and neurosensory abnormalities.Citation3 Hence, DED negatively impacts visual comfort, quality of life, and work productivity. Citation4–6 Tear height meniscus (TMH) is an important clinical marker for DED detection because it helps establish whether the disease is evaporative or aqueous-deficient.Citation7,Citation8 A TMH higher than 0.20 mm is considered normal, while lower values indicate an aqueous-deficient DED.Citation7 Furthermore, TMH correlates well with symptoms and tear function tests and is a useful therapeutic effect indicator of artificial tears in DED.Citation9,Citation10

TMH is affected by factors such as age, ethnicity,Citation7 low relative humidity and high air velocity,Citation11 contact lenses wear, eye drops instillation,Citation12 tear osmolarity,Citation13 time after blinking, diurnal variations, and conjunctival folds.Citation14 TMH is clinically evaluated through slit-lamp assessment, either by comparison with a variable-height beam of light,Citation7 or with a calibrated graticule into the slit-lamp eyepiece.Citation15 Due to its subjective nature, limited image resolution and magnification, slit-lamp assessments are prone to errors.Citation16 Fluorescein sodium can be administered during slit-lamp assessments,Citation17 but it is an invasive procedure that causes reflex tearing, leading to higher inter-observer variability and TMH overestimation.Citation18,Citation19 Some of these drawbacks are avoided with objective and non-invasive techniques, like slit-lamp microscope imaging,Citation20 reflective meniscometry,Citation21 optical coherence tomography (OCT) of the anterior pole,Citation22 and interferometry.Citation5,Citation23 OCT has shown good repeatability in TMH measurement (intraclass correlation coefficient (ICC = 0.900–0.980),Citation14,Citation24 and high sensitivity (67.0–97.4%) and specificity (81.0–90.0%) for the diagnosis of DED.Citation14,Citation25–27 However, the agreement between OCT devices is poor, providing different TMH results.Citation28 Slit-lamp microscope imaging has similar repeatability to the OCT.Citation16 Interferometry is a reliable method for measuring TMH (ICC = 0.870–0.920), with good sensitivity (73%) and specificity (75%) for the diagnosis of DED.Citation16 Overall, recent generation Fourier-domain OCT (FD-OCT) and its two implementations Spectral- and Swept-Source-domain OCT (SD-OCT and SS-OCT) are the more reliable method for TMH measurements,Citation7,Citation29 but are not widely used in clinical practice due to expensive and separate equipment.Citation19

Recently, corneal topographers were incorporated with DED assessment modules that allow non-invasive, and in some cases, automated TMH measurement,Citation8,Citation18 decreasing evaluation time and observer bias.Citation30 Several studies compare different topographers. The Oculus Keratograph 5 M (K5M, OculusOptikgerate GmbH, Wetzlar, Germany) showed good inter-individual variation (ICC = 0.914) and coefficient of variation (CV = 16.4%) as well as good intra-individual variation (ICC = 0.940, CV = 15.9%) for the TMH measurements,Citation1 and good DED diagnosis ability with a cut-off of 0.23 mm (sensitivity of 85.0% and specificity of 77.5%).Citation27 The Oculus Keratograph 5 M measures TMH slightly lower than the FD-OCT.Citation18 In another study, Visionix VX120+ (Visionix-Luneau Technologies, Chartres, France) provided significantly higher TMH measurements than the Medmont E300 (Medmont International Pty Ltd., Melbourne, Australia), with no significant correlations between devices.Citation8 Hence, different topographers are not interchangeable for TMH evaluation, due to different software, hardware, illuminations, and acquisition and measurement protocols.Citation14 Additionally, the high cost and lack of cost-effectiveness studies remain obstacles to corneal topographers widespread clinical use.Citation30 Apart from two studies,Citation14,Citation31 usually only two devices are compared on the same population. Recently, a new corneal topographer, Medmont Meridia (Medmont International Pty Ltd., Victoria, Australia), was proposed, but to the best of our knowledge, no studies have compared it with other devices.

This study aimed to compare TMH measurements between Oculus Keratograph 5 M, Medmont Meridia and the Spectral-domain OCT (Spectralis SD-OCT; Spectralis HRAII-OCT, Heidelberg Engineering GmbH, Heidelberg, Germany) on the same non-clinical participants.

Methods

This was a cross-sectional, observational, and descriptive study with a quantitative methodology approved by the ethics committee of the University of Beira Interior and conducted in accordance with the tenets of the Declaration of Helsinki. Participants were recruited and evaluated at the Experimental and Clinical Centre of Vision Science (CCECV) between March 2023 and May 2023. Informed consent was obtained from all participants prior to data collection.

Participants

The right eyes of 56 participants were included in this study. Each participant underwent an eye examination that included a medical history and monocular and binocular corrected visual acuity (logMAR) at both far (>4 m) and near (40 cm) distances. The anterior pole was evaluated with the Oculus Keratograph 5 M, as it allowed imaging recording.Citation32 The exclusion criteria were previous corneal, refractive, or eyelid surgery; previous ocular trauma, ocular surface disease, or ocular inflammation; diabetes; taking medication that affects the ocular surface and/or tear film; wearing contact lens on the measurement day; use of artificial tears during the previous month; and low quality or uninterpretable corneal topographic and/or SD-OCT images. A sample size of 53 was determined using the G*Power software (version 3.1.9.4), with an effect size of 0.25, significance criteria of α = 0.05, power = 0.80, and a sphericity of 0.75, for the ANOVA repeated measures within factors statistical test.Citation33

Tear meniscus height imaging acquisitions

One well-trained operator conducted image acquisitions on all devices sequentially and randomly using the built-in software of each device. The devices were placed side-by-side in the same room. Image acquisitions were performed between 10 AM and 5 PM diminishing TMH diurnal variation,Citation34 under scotopic conditions to avoid room illumination interference,Citation35 and at the same period to avoid temperature and humidity fluctuations.Citation28 Participants were instructed to remove their glasses (if any), sit up straight, rest their forehead and chin on the corresponding device supports, and to maintain a normal and spontaneous blink to spread the tears evenly and avoid ocular surface dehydration.Citation1 To ensure consistent alignment in the various measurements, participants were instructed to look towards the fixation point of each device. A 3-minute interval between device images was considered, with the participant told to maintain normal blinking while switching between devices. A single image is acquired for each participant at each device. This reflects current clinical practice with limited time per examination. All data was collected at the end of the image acquisitions.

Inter-observer and inter-device analysis of TMH measurements

For inter-observer repeatability, a random selection of participants (n = 20) was considered.Citation1 TMH imaging records were examined by a well-trained observer and a senior optometrist. For the inter-device reproducibility analysis, TMH measurements were performed by a single observer on all participants (n = 56). A single TMH measurement was performed with the device built-in software on the image record as described next.

Oculus Keratograph 5 M

Oculus Keratograph 5 M uses an automated real-time videokeratography imaging method with a non-invasive tear film scanning function.Citation18 Four infrared light-emitting diodes enable imaging under infrared and darker conditions, reducing reflex tearing.Citation1 The built-in software calliper function was used to perform the measurements below the corneal vertex (position specular reflection) at the 6 o’clock position of the cornea (). TMH was defined as the length between the darker edge of the lower lid and the upper border of the reflex line of the tear meniscus.Citation1

Figure 1. Tear meniscus height measurement examples using Oculus Keratograph 5 M (a), Medmont Meridia (b), and Spectralis SD-OCT (c) The images were cropped and resized to enable better visualisation.

Figure 1. Tear meniscus height measurement examples using Oculus Keratograph 5 M (a), Medmont Meridia (b), and Spectralis SD-OCT (c) The images were cropped and resized to enable better visualisation.

Medmont Meridia

The Medmont Meridia provides high-resolution images, with a larger field of view, and DED evaluation protocols. It has a real-time videokeratography imaging system, with a white light placid disk that projects 32 rings on the anterior eye surface, providing a reflection along the tear meniscus border.Citation4,Citation8 TMH measurements were made on the 3 ms captured image of the Non-Invasive Break-Up Time (NIBUT) video built-in evaluation protocol with a 7.0× magnification. This ensures a clear image and avoids prolonged exposure to white light. TMH was defined by the dark edge of the eyelid and the reflected line of the meniscus at the 6 o’clock position of the cornea below the corneal vertex (position specular reflection ()).Citation8 Measurements were performed using the calliper function of the device.

Spectralis SD-OCT

Spectralis SD-OCT uses a super-luminescent diode light source with a wavelength of 870 nm to image the anterior surface of the eye. Scans are performed at 40,000 axial scans per second with an axial resolution of 7 µm, with a transverse resolution of 30 µm,Citation36 and an acquisition time of 12.8 ms.Citation28 Imaging of the TMH was performed using a vertical beam aligned with the corneal vertex (specular reflexion) at the 6 o’clock position of the cornea. TMH measurements were performed using the built-in software calliper function on the first captured image. TMH was defined as the point of touch of the meniscus with the cornea and eyelid ().Citation28,Citation37

Statistical analysis

Statistical analysis was performed with the SPSS software (version 28.0; SPSS Inc. www.ibm.com). Categorical variables were reported as n(%) and continuous variables as mean ± SD (standard deviation) or median and interquartile range. Normality was assessed using the Kolmogorov – Smirnov test. Inter-observer differences were analysed using a paired t-test. Inter-observer repeatability was assessed using within-participant standard deviation (Sw), test–retest repeatability (2.77 Sw), within-participant coefficient of variation (CVws),Citation38 and intraclass coefficient correlation (ICC).Citation39 Inter-device reproducibility was analysed based on the coefficient of variation (CV), intraclass correlation coefficient (ICC), and Pearson correlation coefficient.Citation39 Repeated measures ANOVA with pairwise comparisons was used to determine TMH differences within the imaging devices. Bland-Altman plots were used to evaluate the agreement between the devices. The 95% limits of agreement (LoA) were defined as 1.96×SD of the mean difference. All statistical tests were two-tailed, and the statistical significance was defined as p < 0.05.Citation40

Results

This study enrolled 56 participants (41 females) with a mean age of 43.8 ± 22.4 years (range, 19–88 years). The mean spherical equivalent was −0.77 ± 2.06 D, and the mean distance visual acuity was 0.08 ± 0.20 logMAR. The mean TMH measurement with Oculus Keratograph 5 M, Medmont Meridia and Spectralis SD-OCT were 0.29 ± 0.16 mm, 0.24 ± 0.09 mm and 0.27 ± 0.16 mm, respectively ().

Figure 2. Boxplot showing TMH measurements with different devices. Dashed lines (–) represent median values, grey circles () represent mean values, and plus sign (+) represents outliers.

Figure 2. Boxplot showing TMH measurements with different devices. Dashed lines (–) represent median values, grey circles () represent mean values, and plus sign (+) represents outliers.

shows no inter-observer significant differences (paired t-test, p < 0.001). All the devices achieved an ICC > 0.75 and a CV < 16.53%. TMH measurement comparisons between the three devices () revealed that Medmont Meridia has a lower variability for TMH measurements (CV = 35.75%) than Oculus Keratograph 5 M (CV = 54.46%) and Spectralis SD-OCT (CV = 57.87%). Moderate reliability was found between the three devices (ICC = 0.621, p < 0.001). Repeated measures ANOVA was used to further examine the differences between devices. The sphericity assumption was not verified (Mauchly’s test, XCitation2(2) = 7.189, p = 0.03). A Greenhouse–Geisser correction showed that the TMH measurements between the devices were significantly different (F (1.778, 97.808) = 4.777, p = 0.01). Post-hoc analysis with a Bonferroni adjustment revealed that the TMH measurements from Oculus Keratograph 5 M were significantly higher than those from Medmont Meridia (0.049 (95% CI, 0.01 to 0.09) mm, p = 0.02). TMH measurements obtained by the Spectralis SD-OCT were not statistically significantly different from those obtained by the Oculus Keratograph 5 M (−0.024 (95% CI, −0.06 to 0.01) mm, p = 0.19) and the Medmont Meridia (0.025 (95% CI, −0.02 to 0.07) mm, p = 0.38).

Table 1. Inter-observer repeatability of the TMH by the three devices (n = 20). Mean (M), standard deviation (SD), p-value (p), within-participant coefficient of variation (CVws), within-participant standard deviation (Sw), test–retest repeatability (2.77 Sw), intraclass coefficient correlation (ICC), and Limits of Agreement (LoA).

Table 2. Inter-device reproducibility of the TMH by the three devices (n = 56). Standard deviation (SD), coefficients of variation (CV), intraclass coefficient correlation (ICC), and Limits of agreement (LoA).

The Bland-Altman plots of the difference compared to the mean for the TMH measurements with each device combination are shown in (left column). TMH differences between devices did not fulfil normality (Medmont Meridia – Spectralis SD-OCT, p-value = 0.03 and Oculus Keratograph 5 M – Medmont Meridia, p = 0.03). Since neither a logarithmic transformation nor the ratio of the two TMH values for each participant removes the relationship between the differences and the measurement size, a regression analysis was carried out (, middle column).Citation40 TMH measurements of Oculus Keratograph 5 M were on average (0.05 ± 0.26 mm 95% CI) higher than Medmont Meridia, resulting in higher limits of agreement. This bias is also larger than the mean between Oculus Keratograph 5 M and Spectralis SD-OCT (0.02 ± 0.19 mm 95% CI) and between Medmont Meridia and Spectralis SD-OCT (−0.03 ± 0.25 mm 95% CI).

Figure 3. Bland-Altman plots show differences in TMH measurements for the different devices (n = 56). The left column shows the differences plotted against the mean TMH measurements for each device combination (left). The solid line represents the mean difference. Dashed lines represent 95%LoA (Limits of Agreement). The correlation of the differences with the mean for the TMH measurements is shown in the middle column. Dashed lines represent the regression of 95%LoA. The right column shows the TMH measurement correlation plots for the several devices. The dashed line represents the identity line (y = x).

Figure 3. Bland-Altman plots show differences in TMH measurements for the different devices (n = 56). The left column shows the differences plotted against the mean TMH measurements for each device combination (left). The solid line represents the mean difference. Dashed lines represent 95%LoA (Limits of Agreement). The correlation of the differences with the mean for the TMH measurements is shown in the middle column. Dashed lines represent the regression of 95%LoA. The right column shows the TMH measurement correlation plots for the several devices. The dashed line represents the identity line (y = x).

A positive correlation was found between the mean of TMH values measured with the Oculus Keratograph 5 M and Medmont Meridia, and the difference in the TMH measurements (r2 = 0.62, p < 0.001), indicating that Oculus Keratograph 5 M tends to give higher values as the mean TMH increases. A negative correlation was found between the mean of TMH values measured with the Medmont Meridia and the Spectralis SD-OCT and the difference in the TMH measurements (r2 = -0.59, p < 0.001), indicating that Medmont Meridia tends to give lower values as the mean TMH increases. No significant correlation was found between the mean TMH measured with the Oculus Keratograph 5 M and Spectralis SD-OCT and the difference in the TMH measurements (r2 = 0.05, p = 0.74), suggesting similarity between the two devices. TMH measured with all devices showed a close correlation (r2 = 0.55 to 0.81, p < 0.001) (, right column), with a higher value achieved between Oculus Keratograph 5 M and Spectralis SD-OCT (r2 = 0.81, p < 0.001).

Discussion

This study reports on the use of three different devices to measure tear meniscus height (TMH) in non-clinical participants. Moreover, it is the first time that Medmont Meridia is compared with other devices. All devices achieved a mean TMH higher than 0.20 mm for a non-clinical population. Measurements with the devices are reproducible and repeatable between different observers, but only moderate reliability is found between devices.

The mean TMH with Oculus Keratograph 5 M was 0.29 ± 0.16 mm, Medmont Meridia was 0.24 ± 0.09 mm, and Spectralis SD-OCT was 0.27 ± 0.16 mm. In the literature, TMH measured with the OCT ranged from 0.14 to 0.31 mm in DED participants and 0.21 to 0.37 mm in healthy participants.Citation10,Citation18,Citation25,Citation26,Citation28,Citation37 For the Oculus Keratograph 5 M, TMH measurements ranged from 0.22 to 0.24 mm in DED participants and 0.24 to 0.34 mm in healthy participants.Citation18,Citation28,Citation32,Citation37 No studies have been conducted on the Medmont Meridia, therefore no comparison can be made with it.

In this work, Oculus Keratograph 5 M gives higher values than the Spectralis SD-OCT (an implementation of FD-OCT) with a mean difference of 0.02 mm. This result contradicts previous reports where TMH measured with the Oculus Keratograph 5 M were consistently lower than FD-OCT in both DED and healthy participants.Citation18,Citation28,Citation37 Particularly, Oculus Keratograph 5 M was 0.07 mm lower than the FD-OCT for DED participants,Citation20 0.01 mm for healthy participants,Citation35 and on average 0.03 mm lower in both the DED (p < 0.001) and healthy participants (p = 0.001).Citation37 These findings were justified by the short evaluation time and the infrared illumination of the Oculus Keratograph 5 M that diminishes reflex tearing, compared with the longer adjustment and examination time of FD-OCT devices.Citation18,Citation37 Furthermore, image conversion from the optical to the physical space by OCT algorithms distorts the final sagittal image affecting the TMH measurements.Citation18 The fact that in this work TMH values with the Oculus Keratograph 5 M are higher than Spectralis SD-OCT may be explained by the considered non-clinical participants. The literature describes three groups of participants, namely, DED,Citation18,Citation32,Citation37 healthy,Citation10,Citation28,Citation32,Citation37 and non-clinical.Citation8,Citation14,Citation31 DED participants have symptoms and/or signs of dry eye. Healthy (control or normal) participants do not have symptoms and/or signs related to DED. Non-clinical participants are not subjected to any inclusion or exclusion criteria based on the presence or absence of symptoms and/or signs of dry eye. Hence, a non-clinical sample has an unknown ratio of DED and healthy participants. Nevertheless, our results indicate that the three devices offer different results for the same participants. It should be noted that for non-clinical participants, Oculus Keratograph 5 M reached a TMH mean value of 0.29 ± 0.08 mm which is remarkably consistent with our study (0.29 ± 0.16 mm).Citation31

For the Medmont Meridia, only a mean difference of 0.03 mm with Spectralis SD-OCT and 0.05 mm with the Oculus Keratograph 5 M was found. No study has considered the Medmont Meridia. The closest one compares the Medmont E300 (Medmont Meridia predecessor) with another Placido disk-based topographer (Visionix VX120+) on non-clinical participants.Citation8 The TMH with the Visionix VX120+ was found to be higher (about 0.03 mm) than with the Medmont E300. This finding and this study results suggest that differences between the Medmont Meridia and other devices may be clinically insignificant.

Inter-observer analysis results show that TMH measured with the Oculus Keratograph 5 M, Medmont Meridia and the Spectralis SD-OCT are repeatable and reproducible at a similar level. In terms of reliability, a good ICC (≥0.75),Citation32 was found for all devices (≥0877), with the Spectralis SD-OCT being the highest (0.989). The Spectralis SD-OCT also showed the best repeatability (CV = 2.87%) followed by the Oculus Keratograph 5 M (CV = 11.08%) and the Medmont Meridia (CV = 16.53%). Several results have been published regarding the inter-observer reproducibility and reliability in DED and healthy participants.Citation18,Citation28,Citation32,Citation37 For DED participants, it was found good reliability for Oculus Keratograph 5 M (ICC = 0.985) and the SD-OCT (ICC = 0.993), as well as good reproducibility (CV of 5.58% and 5.05% for Oculus Keratograph 5 M and SD-OCT respectively).Citation18 In contrast, for healthy participants it was found a good reliability (ICC) in the range of 0.700–0.760 for the Oculus Keratograph 5 M and 0.920–0.930 for the SD-OCT,Citation35 which is remarkably similar to our study. The reproducibility (CV) ranged from 24.0% to 19.0% for the Oculus Keratograph 5 M and 11.0% to 12.0% for the SD-OCT. Furthermore, in healthy participants, it was achieved good reliability (ICC = 0.914), as well as good repeatability (CV = 16.40%).Citation37 In another study, the Oculus Keratograph 5 M presented good repeatability for both DED participants (CV = 16.08%) and healthy participants (CV = 19.76%).Citation32 Although there are no studies with the Medmont Meridia our results suggest that this device is very similar to Oculus Keratograph 5 M. Overall and despite the fact that our findings are for non-clinical participants, they are in good agreement with the literature in considering the SD-OCT as the device with the best inter-observer reliability and reproducibility,Citation28,Citation37 but closely followed by the Oculus Keratograph 5 M and the Medmont Meridia.

Inter-device analysis showed only moderate reliability between the devices (ICC = 0.621). An ANOVA with repeated measures and Bland-Altman plots analysis revealed that only the Spectralis SD-OCT and Oculus Keratograph 5 M yielded similar results. Particularly, Oculus Keratograph 5 M gives higher TMH values than Medmont Meridia, but TMH measurements with Spectralis SD-OCT do not statistically differ from either Oculus Keratograph 5 M or Medmont Meridia, but Medmont Meridia tends to give lower values than Spectralis SD-OCT. The Oculus Keratograph 5 M and Spectralis SD-OCT exhibited larger correlation between TMH measurements (r2 = 0.81, p < 0.001). Several studies compare the Oculus Keratograph 5 M and the FD-OCT with conflicting results.Citation18,Citation28,Citation37 For DED participants, it was found that TMH measurements with the Oculus Keratograph 5 M closely correlated with the FD-OCT, with good repeatability and reliability, although Oculus Keratograph 5 M tended to give lower values at higher TMH.Citation18 For healthy participants, no significant differences between the two devices were found, although Bland-Altman plots revealed a poor agreement.Citation28 To explain the conflicting results, another study considered both healthy and DED participants and found a poor agreement between the devices in both groups, with FD-OCT showing better repeatability than the Keratograph 5 M in both groups.Citation37 Hence, the literature consistently reports no agreement between TMH measurements with different devices.Citation8 Various reasons are given for this lack of agreement. Although recent corneal topographers are high-resolution devices,Citation8,Citation18 meniscus delimitation is difficult to perceive due to image artefacts caused by eyelash shadows and reflexes. Furthermore, all the devices have their own proprietary software for image processing and TMH measurements. Since neither device automatically delineates the upper lid margin of the lower meniscus, differences between TMH measurement protocols are unclear.Citation8,Citation28 Another factor is the imaging and illumination method of each device. Medmont Meridia uses a white light Placido disk with 32 rings, while Oculus Keratograph 5 M captures images under infrared conditions. Spectralis SD-OCT uses a super-luminescent diode light source to image the eye anterior surface.Citation36 The influence of each illumination system is unclear and may lead to variability and poor agreement between TMH measurements.Citation8

Our study is not without some limitations. Although the reduced number of participants (n = 56) may limit the comparison between devices, it is in the range of previous studies (22 to 66),Citation28,Citation31 supporting the results’ robustness. The considered young and non-clinical participants do not allow the separation of healthy and DED participants and DED diagnosis evaluation. Since DED is affected by age, sex, and ethnicity, further studies must consider both healthy and DED participants across these groups. Moreover, since devices are not interchangeable, future works must define cut-off values for the previous described populations and compare TMH measurements of aqueous and lipidic tears and the diagnosis of evaporative and aqueous-deficient DED. As in a previous study,Citation8 only a single measurement per participant and device was considered. Although this seems to be a limitation due to the dynamic nature of the tear film,Citation28 it is a common clinical practice due to the limited time per patient. To ensure that measurements were taken as closely as possible in the same eyelid position between the devices, participants were instructed to look at the fixation point given by the device, as this is usually aligned with the primary gaze position. Measurements were made below the corneal vertex at the cornea 6 o’clock position with the lower eyelid.Citation28 Furthermore, the repeatability of all the devices is well established in the literature,Citation8,Citation28 and since our measurements were performed randomly and sequentially in the same period of the day, the influence of a single measure is minimised. The simultaneous analysis of three distinct devices in the same population is a significant advantage of our study. Despite the use of non-clinical participants, the obtained results remain valid.

Conclusion

For non-clinical participants, Spectralis SD-OCT provides better inter-observer repeatability and reproducibility. Medmont Meridia and Oculus Keratograph 5 M cannot be used interchangeably in clinical practice. Spectralis SD-OCT exhibited strong correlation and good agreement with Oculus Keratograph 5 M. Overall, our results indicate that care must be taken when using these devices interchangeably in clinical practice. Additionally, every device measures TMH differently, suggesting that more studies are needed to determine individual device cut-off DED values by age and sex.

Acknowledgements

The authors would like to thank Optometron: Equipamentos Técnicos E Electrónicos Lda for providing the Oculus Keratograph 5 M, Optiforum: Contactologia Avançada for providing the Medmont Meridia, and the CCECV for providing the Spectralis SD-OCT.

Disclosure statement

No potential conflict of interest was reported by the author(s).

References

  • Wei A, Le Q, Hong J et al. Assessment of lower tear meniscus. Optom Vis Sci 2016; 93: 1420–1425. doi:10.1097/OPX.0000000000000986.
  • Stapleton F, Alves M, Bunya VY et al. TFOS DEWS II epidemiology report. Ocul Surf 2017; 15: 334–365. doi:10.1016/j.jtos.2017.05.003.
  • Craig JP, Nichols KK, Akpek EK et al. TFOS DEWS II definition and classification report. Ocul Surf 2017; 15: 276–283. doi:10.1016/j.jtos.2017.05.008.
  • Lim J, Wang MTM, Craig JP. Evaluating the diagnostic ability of two automated non-invasive tear film stability measurement techniques. Cont Lens Anterior Eye 2021; 44: 101–362. doi:10.1016/j.clae.2020.08.006.
  • Herbaut A, Liang H, Denoyer A et al. Analyse du film lacrymal et évaluation de sa qualité optique : une revue de la littérature. J Fr Ophtalmol 2019; 42: 226–243. doi:10.1016/j.jfo.2018.10.004.
  • Wang MTM, Craig JP. Natural history of dry eye disease: perspectives from inter-ethnic comparison studies. Ocul Surf 2019; 17: 424–433. doi:10.1016/j.jtos.2019.03.004.
  • Wolffsohn JS, Arita R, Chalmers R et al. TFOS DEWS II diagnostic methodology report. Ocul Surf 2017; 15: 539–574. doi:10.1016/j.jtos.2017.05.001.
  • Martínez-Plaza E, Molina-Martín A, Piñero DP. Agreement of tear break-up time and meniscus height between medmont e300 and visionix VX120+. Appl Sci 2022; 12: 45–89. doi:10.3390/app12094589.
  • Li J, Shen M, Wang J et al. Clinical significance of tear menisci in dry eye. Eye Contact Lens Sci Clin Pract 2012; 38: 183–187. doi:10.1097/ICL.0b013e318252ce0c.
  • Lam AK, Tai SK, Chan JK et al. Lower tear meniscus height measurements using keratography and swept-source optical coherence tomography and effect of fluorescein instillation methods. Curr Eye Res 2019; 44: 1203–1208. doi:10.1080/02713683.2019.1629598.
  • Koh S, Tung C, Kottaiyan R et al. Effect of airflow exposure on the tear meniscus. J Ophthalmol 2012; 2012: 1–6. doi:10.1155/2012/983182.
  • Nagahara Y, Koh S, Maeda N et al. Prominent decrease of tear meniscus height with contact lens wear and efficacy of eye drop instillation. Eye Contact Lens Sci Clin Pract 2015; 41: 318–322. doi:10.1097/ICL.0000000000000134.
  • García-Resúa C, Pena-Verdeal H, Remeseiro B et al. Correlation between tear osmolarity and tear meniscus. Optom Vis Sci 2014; 91: 1419–1429. doi:10.1097/OPX.0000000000000412.
  • Niedernolte B, Trunk L, Wolffsohn JS et al. Evaluation of tear meniscus height using different clinical methods. Clin Exp Optom 2021; 104: 583–588. doi:10.1080/08164622.2021.1878854.
  • Imamura H, Tabuchi H, Nakakura S et al. Usability and reproducibility of tear meniscus values generated via swept-source optical coherence tomography and the slit lamp with a graticule method. Int Ophthalmol 2018; 38: 679–686. doi:10.1007/s10792-017-0517-3.
  • Santodomingo-Rubido J, Wolffsohn JS, Gilmartin B. Comparison between graticule and image capture assessment of lower tear film meniscus height. Cont Lens Anterior Eye 2006; 29: 169–173. doi:10.1016/j.clae.2006.05.004.
  • Kawai M, Yamada M, Kawashima M et al. Quantitative evaluation of tear meniscus height from fluorescein photographs. Cornea 2007; 26: 403–406. doi:10.1097/ICO.0b013e318033c242.
  • Baek J, Doh SH, Chung SK. Comparison of tear meniscus height measurements obtained with the keratograph and Fourier domain optical coherence tomography in dry eye. Cornea 2015; 34: 1209–1213. doi:10.1097/ICO.0000000000000575.
  • Bandlitz S, Purslow C, Murphy PJ et al. Comparison of a new portable digital meniscometer and optical coherence tomography in tear meniscus radius measurement. Acta Ophthalmol 2014; 92: 112–118. doi:10.1111/aos.12275.
  • Yang J, Zhu X, Liu Y et al. TMIS: a new image‐based software application for the measurement of tear meniscus height. Acta Ophthalmol 2019; 97: 973–980. doi:10.1111/aos.14107.
  • Yokoi N, Bron A, Tiffany J et al. Reflective meniscometry: a non-invasive method to measure tear meniscus curvature. Br J Ophthalmol 1999; 83: 92–97. doi:10.1136/bjo.83.1.92.
  • Binotti WW, Bayraktutar B, Ozmen MC et al. A review of imaging biomarkers of the ocular surface. Eye Contact Lens Sci Clin Pract 2020; 46: S84–105. doi:10.1097/ICL.0000000000000684.
  • Fodor E, Hagyó K, Resch M et al. Comparison of tearscope-plus versus slit lamp measurements of inferior tear meniscus height in normal individuals. Eur J Ophthalmol 2010; 20: 819–824. doi:10.1177/112067211002000502.
  • Canan H, Altan-Yaycioglu R, Ulas B et al. Interexaminer reproducibility of optical coherence tomography for measuring the tear film meniscus. Curr Eye Res 2014; 39: 1145–1150. doi:10.3109/02713683.2014.898311.
  • Akiyama R, Usui T, Yamagami S. Diagnosis of dry eye by tear meniscus measurements using anterior segment swept source optical coherence tomography. Cornea [Internet] 2015; 34: S115–20. doi:10.1097/ICO.0000000000000583.
  • Shinzawa M, Dogru M, Miyasaka K et al. Application of CASIA SS-1000 optical coherence tomography tear meniscus imaging in testing the efficacy of new strip meniscometry in dry eye diagnosis. Eye Contact Lens Sci Clin Pract 2018; 44: S44–9. doi:10.1097/ICL.0000000000000312.
  • Wang Y, Xu Z, Gong Q et al. The role of different tear volume detection methods in the evaluation and diagnosis of mild dry eye disease. Transl Vis Sci Technol 2022; 11: 15. doi:10.1167/tvst.11.3.15.
  • Arriola-Villalobos P, Fernández-Vigo JI, Díaz-Valle D et al. Assessment of lower tear meniscus measurements obtained with keratograph and agreement with Fourier-domain optical-coherence tomography. Br J Ophthalmol 2015; 99: 1120–1125. doi:10.1136/bjophthalmol-2014-306453.
  • de Boer JF, Leitgeb R, Wojtkowski M. Twenty-five years of optical coherence tomography: the paradigm shift in sensitivity and speed provided by Fourier domain OCT [Invited]. Biomed Opt Express 2017 Jun 15; 8: 3248–3280. doi:10.1364/BOE.8.003248.
  • Di Cello L, Pellegrini M, Vagge A et al. Advances in the noninvasive diagnosis of dry eye disease. Appl Sci 2021; 11: 10384. doi:10.3390/app112110384.
  • Garcia-Terraza AL, Jimenez-Collado D, Sanchez-Sanoja F et al. Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface. Front Med 2022; 9: 1–7. doi:10.3389/fmed.2022.893688.
  • Tian L, Qu J, Zhang X et al. Repeatability and reproducibility of noninvasive keratograph 5m measurements in patients with dry eye disease. J Ophthalmol 2016; 2016: 1–6. doi:10.1155/2016/8013621.
  • Faul F, Erdfelder E, Lang A-G et al. G*power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 2007; 39: 175–191. doi:10.3758/BF03193146.
  • Shen M, Wang J, Tao A et al. Diurnal variation of upper and lower tear menisci. Am J Ophthalmol 2008; 145: 801–806. doi:10.1016/j.ajo.2007.12.024.
  • Doughty MJ, Laiquzzaman M, Button NF. Video-assessment of tear meniscus height in elderly caucasians and its relationship to the exposed ocular surface. Curr Eye Res 2001; 22: 420–426. doi:10.1076/ceyr.22.6.420.5487.
  • Kiernan DF, Mieler WF, Hariprasad SM. Spectral-domain optical coherence tomography: a comparison of modern high-resolution retinal Imaging Systems. Am J Ophthalmol 2010; 149: 18–31. doi:10.1016/j.ajo.2009.08.037.
  • Chen M, Wei A, Xu J et al. Application of keratograph and Fourier-domain optical coherence tomography in measurements of tear meniscus height. J Clin Med 2022; 11: 13–43. doi:10.3390/jcm11051343.
  • Bland JM, Altman DG. Statistics notes: measurement error proportional to the mean. BMJ Publ Gr 1996; 313: 106–106. doi:10.1136/bmj.313.7049.106.
  • Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016; 15: 155–163. doi:10.1016/j.jcm.2016.02.012.
  • Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135–160. doi:10.1177/096228029900800204.